.General Introduction

Yokogawa ZR202S zirconia oxygen analyzer is the integrated explosion-proof type including both probe detector and converter that are used for hazardous zone Exd. It does not need a sampling device, and allow direct installation of the probe in the wall of a flue or furnace to measure the concentration of oxygen in the stack gas. The converter displays the cell temperature and cell emf in addition to the oxygen concentration which is most suitable for monitoring combustion and controlling the low-oxygen combustion of various industrial furnaces in explosive atmosphere at petroleum refinery, petrochemical plant, and natural gas plant.

2.General Specification

Main Characters

Description

Sample Gas Temperature

0 to 700℃

Sample Gas Pressure

-5 to + 250kPa

Probe Length

 0.4, 0.7, 1.0, 1.5, 2.0m

Probe Material

SUS316 (JIS)

Ambient Temperature

-20 to +55℃

Reference Gas System

Instrument Air

Instrument Air System

50kPa + the pressure inside the furnace

Wetted Material

SUS316 (JIS), Zirconia, SUS304 (JIS) or   ASTM grade 304 (flange), Hastelloy B, Inconel 600, 601)

Gas Connection:

Rc1/4 or 1/4 NPT (Female)

Wiring Connection

M20 × 1.5, 1/2 NPT

Installation

Flange mounting

Probe Mounting Angle

Horizontal to vertically downward.

Output

1*4-20ma Hart, 2* contact signal

Display

6-digit LCD


3.Model Selection

Model

Suffix Codes

Description

ZR202S












Integrated type   Explosion-proof Zirconia Oxygen Analyzer

Explosion proof Approval

-A











ATEX certified flameproof

-B











FM certified explosion-proof

-C











CSA certified explosion-proof

-D











IECEx certified flameproof

-Q











EAC with PA certified explosion-proof

-R











EAC certified explosion-proof

-K











KOSHA certified explosion-proof

Length


-040










0.4 m


-070










0.7 m


-100










1.0 m


-150










1.5 m


-200










2.0 m

Wetted material



-S









Stainless steel



-C









Stainless steel with Inconel calibration gas tube

Flange




-A








ANSI Class 150 2 RF




-B








ANSI Class 150 3 RF




-C








ANSI Class 150 4 RF




-E








DIN PN10 DN50 A




-F








DIN PN10 DN80 A




-G








DIN PN10 DN100 A




-K








JIS 5K 65 FF




-L








JIS 10K 65 FF




-M








JIS 10K 80 FF




-P








JIS 10K 100 FF




-Q








JIS 5K 32 FF (for high temperature use)




-R








JPI Class 150 4 RF




-S








JPI Class 150 3 RF




-W








Westinghouse

Auto Calibration





-N







Not required





-A







Horizontal mounting





-B







Vertical mounting

Reference gas






-E






External connection (Instrument air)

Gas Thread







-R





Rc1/4







-T





1/4NPT(Female)

Connection box thread








-M




M20 x1.5








-T




1/2 NPT

Instruction manual









-E



English









-K



Korea

--










-A


Always -A

Options











/C

Inconel bolt











/CV

Check valve











/SV

Stop valve











/H

Hood











/SCT

Stainless steel tag plate











/PT

Printed tag plate











/C2

NAMUR NE43 Failure alarm down-scale











/C3

NAMUR NE43 Failure alarm up-scale


4.Heater Assembly

Model

Suffix Codes

Description

ZR202A




Heater Assembly for ZR202S

Length

-040



0.4 m

-070



0.7 m

-100



1.0 m

-150



1.5 m

-200



2.0 m

Jig for change


-A


with Jig


-N


None

-



-A

Always A